The Kahook Dual Blade (KDB) is a modern goniotomy device. The goal for inventing the KDB device was to find a better and more efficient method for removing a complete strip of trabecular meshwork (TM) using an ab interno approach while minimizing damage to surrounding tissues.
CHECK LINK: https://ourgsc.blogspot.com/search?q=kahook
The iStent is a minimally invasive glaucoma
surgery (MIGS) device devised to bypass the obstructed TM and allow the flow of aqueous from the anterior chamber into the Schlemm’s
canal.
CHECK LINK: https://ourgsc.blogspot.com/search?q=iStent
A systematic review and meta-analysis
comparing the KDB with iStent implantation with phacoemulsification was done by
Guedes et al.
Fourteen studies were included with a total
of 1959 eyes (958 phaco-KDB, and 1000 phaco-Stent including 753 phaco-iStent
and 207 phaco-iStent inject).
The combined findings showed significantly
higher rates of surgical success in the phaco-KDB group versus the phaco-Stent
group (odds ratio: 0.68; 95% CI: 0.50 to 0.92; P = 0.01; I2 = 40%), and greater
IOP reduction in the phaco-KDB group versus the phaco-Stent group at month 6
(MD: 1.13 mm Hg; 95% CI: 0.43 to 1.83; P = 0.002; I2 = 51%).
By month 12, both groups demonstrated
similar IOP reduction. Subgroup analysis at month 12 showed greater IOP
reduction in the phaco-KDB group versus the phaco-iStent group (MD: 1.69 mm Hg;
95% CI: 0.44 to 2.95; P = 0.008; I2 = 74%). However, compared with the phaco
iStent inject group, there was no significant difference in IOP reduction (MD:
−0.72 mm Hg; 95% CI: −3.69 to 2.24; P = 0.63; I2 = 64%). Medication reduction
and the incidence of adverse events were comparable between groups.
RESULT:
The study concluded that KDB goniotomy may offer better surgical success compared with Stent implantation when used in combination with phacoemulsification. KDB goniotomy demonstrated better IOP reduction at month 12 compared with iStent. However, iStent inject showed a similar IOP reduction to KDB.
REFERENCE:
Guedes J, Amaral DC, de Oliveira Caneca K,
Cassins Aguiar EH, de Oliveira LN, Mora-Paez DJ, Cyrino LG, Louzada RN, Moster
MR, Myers JS, Schuman JS, Shukla AG, Shalaby WS. Kahook Dual Blade Goniotomy
Versus iStent Implantation Combined With Phacoemulsification: A Systematic
Review and Meta-Analysis. J Glaucoma. 2025 Mar 1;34(3):232-247. doi:
10.1097/IJG.0000000000002522. Epub 2024 Dec 9. PMID: 39641580.