Sunday, November 22, 2026

TONOMETRY ERRORS FROM FLUORESCEIN

 


In Goldmann Applanation Tonometry (GAT), the amount of fluorescein directly alters the width of the semicircular mires, changing the appearance where the inner edges appear to meet.

With excessive fluorescein, the mires appear as thick, wide, and overly bright semicircles. The wide mires displace the inner edges inward. It takes extra force from the tonometer dial to push the thick rings apart until their inner edges touch. Therefore, the IOP is measured as falsely high.

With insufficient fluorescein, thin, faint, or broken semicircles are seen. The true outer apex of the tear meniscus is not clearly visible. The inner edges appear to touch before the full 3.06 mm corneal area is actually flattened. Therefore, the IOP is measured as falsely low.

Ideally, the observer should identify crisp, thin semicircles (approx. 1/10th the width of the mire diameter) to get the correct IOP.




Astigmatism, if significant, may give distorted mires as well as leading to mechanically induced errors. If over 3 dioptres, the average reading of two can be taken with the prism rotated 90° for the second, or optimally the prism is rotated so that the red line on the tonometer housing is aligned with the prescription of the minus axis.




How to Correct Measurement Errors

  • If mires are too thick. Ask the patient to blink a few times, gently blot the closed eyelids or tear film at the lateral canthus with a clean tissue, and wipe the tonometer prism tip dry before re-measuring.
  • If mires are too thin. Instill a tiny additional amount of fluorescein using a moistened strip (shaking off excess fluid first) or a diluted anesthetic-fluorescein drop, wait a few seconds for tear film stabilization, and repeat the measurement.




TONOMETRY ERRORS FROM FLUORESCEIN

  In Goldmann Applanation Tonometry (GAT), the amount of fluorescein directly alters the width of the semicircular mires, changing the appea...